Device for controlling the pH of the reaction tank. For pH adjustment of the reaction solution. Visualization possible on a computer. Videos are currently being streamed on our website, and demos are available (Kanto and Chubu regions)!
Acid and alkaline solutions are discharged into the reaction tank as droplets based on pH values, controlled by a combination of an air pump and an electromagnetic valve. The pH is measured with a pH electrode, and based on this pH, a specified amount of liquid is discharged to automatically adjust the pH. The minimum discharge volume is 10 µl, allowing for precise pH control of fine liquids. Control with a narrow nozzle is optimal for the pretreatment (neutralization) of environmental samples. The adjustment range is pH 0-14. The standard pH measurement unit is HORIBA's D-72, making it ideal for medium adjustment and reaction liquid control. By using optional software, pH changes can be easily "visualized." The system is provided, including the additive liquid container. It features remote control with a simple titration function, making it perfect for understanding the properties of solutions with unknown concentrations.
For pH control of plating solutions, the previously complicated pH adjustments and reaction tank controls are now automated under computer control through valve opening and closing, simplifying solution adjustments and reaction control. The pH adjustment features include:
- Titration mode
- Mode to maintain pH at a constant level over time
- Function to gradually change controlled pH (only during remote control)
It is packed with features. Customization can be done to fit your experiments and protocols. Demonstrations are currently being accepted in the Kanto-Chubu region!!!